The Gap (Mount Isa) 2019 hourly air quality...

Air Quality Monitoring - 2019

Created 05/05/2025

Updated 05/05/2025

Period of operation: 1 January to 31 December 2019
Site location: Latitude: -20.7264; Longitude: 139.4977
AS/NZS 3580.1.1 site classification: Neighbourhood
AS/NZS 3580.1.1 compliance (pollutants): Complies fully
AS/NZS 3580.14 compliance (meteorology): Complies, except for:

  • Buildings higher than wind sensor located within 20 metres north east and east of the monitoring site.
  • Monitoring shelter located on concreted surface.

Wind: Ultrasonic sensor

  • Height above ground level: 10 metres.
  • The wind direction sensor is aligned to magnetic north. The reported wind direction is referenced to true north by application of a magnetic declination correction of +6 degrees to the sensor output value.

Temperature: Capacitive ceramic sensor

  • Height above ground level: 3 metres.

Sulfur dioxide: Ultraviolet fluorescence analyser operated in accordance with AS 3580.4.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

PM10: Tapered Element Oscillating Balance (TEOM) Model 1405 operated in accordance with AS/NZS 3580.9.16 (excluding FDMS requirements)

  • Measurement values do not compensate for loss of semi-volatile components from the collected particulate matter.
  • Reported concentrations are uncorrected instrument output values.
  • Negative hourly PM10 concentrations less than -0.5 µg/m^3 are removed and concentrations between -0.5 µg/m^3 and 0.0 µg/m^3 are set to 0.0 µg/m^3.

Source

This data file or API can be downloaded from:

https://www.data.qld.gov.au/dataset/827a95bc-7160-4510-9a75-b4b4d22042a6/resource/59138e25-d058-40af-92ab-dc395a3048de/download/thegap-aq-2019.csv