Rocklea (South East Queensland) 2021 hourly...

Air Quality Monitoring - 2021

Created 05/05/2025

Updated 05/05/2025

Period of operation: 1 January to 31 December 2021
Site location: Latitude: -27.5358; Longitude: 152.9934
AS/NZS 3580.1.1 site classification: Neighbourhood
AS/NZS 3580.1.1 compliance (pollutants): Complies, except for:

  • Isolated trees above height of inlet within 10 metres west of the monitoring site.

AS/NZS 3580.14 compliance (meteorology): Complies, except for:

  • Isolated trees within 10 metres west of the monitoring site.
  • Temperature and relative humidity sensors located 10 metres above ground level.
  • Rainfall measured by acoustic sensor.

Wind: Ultrasonic sensor

  • Height above ground level: 10 metres.
  • The wind direction sensor is aligned to magnetic north. The reported wind direction is referenced to true north by application of a magnetic declination correction of +11 degrees to the sensor output value.

Temperature: Capacitive ceramic sensor

  • Height above ground level: 10 metres.

Relative humidity: Capacitive thin film polymer sensor

  • Height above ground level: 10 metres.

Rainfall: Vaisala RAINCAP® precipitation sensor.

  • Height above ground level: 10 metres.

Barometric pressure: Capacitive silicon sensor.

  • Height above ground level: 10 metres.

Solar radiation: Pyranometer
Ozone: UV absorption analyser operated in accordance with AS/NZS 3580.6.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

Oxides of nitrogen: Chemiluminescence analyser operated in accordance with AS/NZS 3580.5.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

PM10/PM2.5: Dichotomous Tapered Element Oscillating Balance (TEOM) Model 1405-DF fitted with Filter Dynamics Measurement System (FDMS) operated in accordance with AS/NZS 3580.9.13

  • The FDMS system compensates for the loss of semi-volatile components from the collected particulate matter.
  • Reported concentrations are uncorrected instrument output values.
  • The reported PM10 and PM2.5 concentrations are calculated from running 1-hour average concentrations updated at six minute intervals.
  • Negative hourly PM10/PM2.5 concentrations down to -5 µg/m^3 resulting from instrument noise at low particle concentrations are reported.

Visibility-reducing particles: Nephelometer operated in accordance with AS/NZS 3580.12.1

  • Detector spectral response centred at 520 nm.
  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.