Wynnum North (South East Queensland) 2016...

Air Quality Monitoring - 2016

Created 05/05/2025

Updated 20/05/2025

Industry-operated monitoring site: Data presented as received by Queensland Government
Period of operation: 1 January to 31 December 2016
Site location: Latitude: -27.4296; Longitude: 153.1581
AS/NZS 3580.1.1 site classification: Neighbourhood
AS/NZS 3580.1.1 compliance (pollutants): Complies fully
AS/NZS 3580.14 compliance (meteorology): Complies fully

  • Wind, temperature, relative humidity, rainfall and barometric pressure sensors located 10 metres above ground level.

Wind sensor: Ultrasonic sensor
Temperature: Capacitive ceramic sensor
Relative humidity: Capacitive thin film polymer sensor
Rainfall: Vaisala RAINCAP® precipitation sensor.
Barometric pressure: Capacitive silicon sensor.
Oxides of nitrogen: Chemiluminescence analyser operated in accordance with AS/NZS 3580.5.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

Sulfur dioxide: Ultraviolet fluorescence analyser operated in accordance with AS 3580.4.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

PM10/PM2.5: Dichotomous Tapered Element Oscillating Balance (TEOM) Model 1405-D operated in accordance with AS/NZS 3580.9.13 (excluding FDMS requirements)

  • Measurement values do not compensate for loss of semi-volatile components from the collected particulate matter.
  • Reported concentrations are uncorrected instrument output values.
  • Negative hourly PM10/PM2.5 concentrations down to -5 µg/m^3 resulting from instrument noise at low particle concentrations are reported.

Source

This data file or API can be downloaded from:

https://www.data.qld.gov.au/dataset/fc8d4550-1d2f-4e56-a2db-7b5f11b903ed/resource/2dc1cbd8-0f7c-4094-9f05-c8603c8fc0fd/download/wynnumnorthaq2016.csv