Springwood (South East Queensland) 2020 hourly...

Air Quality Monitoring - 2020

Created 24/02/2025

Updated 24/02/2025

Period of operation: 1 January to 31 December 2020
Site location: Latitude: -27.6125; Longitude: 153.1359
AS/NZS 3580.1.1 site classification: Neighbourhood
AS/NZS 3580.1.1 compliance (pollutants): Complies, except for:

  • Trees above height of PM10/PM2.5 inlets within 5 metres south and west of the monitoring site.

AS/NZS 3580.14 compliance (meteorology): Complies, except for:

  • Located below the upper level of a ridge running roughly northwest-southeast.
  • Trees within 10 metres of the monitoring site (below height of wind sensor).

Wind: Ultrasonic sensor

  • Height above ground level: 10 metres.
  • The wind direction sensor is aligned to magnetic north. The reported wind direction is referenced to true north by application of a magnetic declination correction of +11 degrees to the sensor output value.

Temperature: Capacitive ceramic sensor

  • Height above ground level: 3 metres.

Relative humidity: Capacitive thin film polymer sensor

  • Height above ground level: 3 metres.

Barometric pressure: Capacitive silicon sensor

  • Height above ground level: 3 metres.

Ozone, nitric oxide, nitrogen dioxide, sulfur dioxide, benzene, toluene, xylenes (total) and formaldehyde (until 8 May): Differential Optical Absorption Spectroscopy (DOAS) analyser operated in accordance with AS/NZS 3580.15

  • Values are average concentrations over 250 metre light path running east-west from (-27.6117, 153.1382) to (-27.6125, 153.1359).
  • Instrument output offset and/or scaled based on instrument calibration data.

Ozone (from 13 May): UV absorption analyser operated in accordance with AS/NZS 3580.6.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

Oxides of nitrogen (from 13 May): Chemiluminescence analyser operated in accordance with AS/NZS 3580.5.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

Sulfur dioxide (from 13 May): Ultraviolet fluorescence analyser operated in accordance with AS 3580.4.1

  • Instrument output offset and/or scaled based on calibration and daily zero/span response data.

PM10/PM2.5: Dichotomous Tapered Element Oscillating Balance (TEOM) Model 1405-DF fitted with Filter Dynamics Measurement System (FDMS) operated in accordance with AS/NZS 3580.9.13

  • The FDMS system compensates for the loss of semi-volatile components from the collected particulate matter.
  • Reported concentrations are uncorrected instrument output values.
  • The reported PM10 and PM2.5 concentrations are calculated from running 1-hour average concentrations updated at six minute intervals.
  • Negative hourly PM10/PM2.5 concentrations down to -5 µg/m^3 resulting from instrument noise at low particle concentrations are reported.

Source

This data file or API can be downloaded from:

https://www.data.qld.gov.au/dataset/e732d13e-6be1-4f69-9266-db25998e9f2c/resource/78fd3d31-dbd3-484c-8a85-042cbf25987c/download/springwood-aq-2020.csv